Electromagnetic compatibility (EMC)
Remark:
The Product subsequently referred to herein always denotes the Vacuson 18 / 40 / 60.
Changes or modifications to this product not expressly approved by the manufacturer may result in increased emissions or decreased immunity
performance of the product and could cause EMC issues with this or other equipment. This product is designed and tested to comply with
applicable regulations regarding EMC and shall be installed and put into service according to the EMC information stated as follows.
WARNING
Use of portable phones or other radio frequency (RF) emitting equipment, including accessories (antennas e.g.) in distances below
30 cm (12 inches) to the product, may cause unexpected or adverse operation.
WARNING
The product shall not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, the product
shall be tested to verify normal operation in the configuration in which it is being used.
Compliant Cables and Accessories
WARNING
The use of accessories, transducers and cables other than those specified may result in increased emissions or decreased immunity
performance of the product.
Guidance and manufacturer's declaration – electromagnetic emissions
The Product is intended for use in the electromagnetic environment specified below. The customer or the user of the Product should
assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Product is intended for use in the electromagnetic environment specified below. The customer or the user of the Product should
assure that it is used in such an environment.
Immunity tests
Electrostatic discharge
+/- 6 kV contact
(ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power supply
transient/burst
lines
IEC 61000-4-4
+/- 1 kV for input/output
lines
Surge
+/- 1 kV differential mode
IEC 61000-4-5
+/- 2 kV common mode
Voltage dips, short
< 5 % U
interruptions and voltage
(> 95 % dip in U
variations on power
for 0,5 cycle
supply input lines
40 % U
IEC 61000-4-11
(60 % dip in U
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
< 5 % U
(> 95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60Hz) magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
complies
IEC 60601
Compliance level
Test level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 1 kV differential mode
+/- 2 kV common mode
< 5 % U
T
T
)
(> 95 % dip in U
T
for 0,5 cycle
40 % U
T
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
T
)
(30 % dip in U
T
for 25 cycles
< 5 % U
T
T
)
(> 95 % dip in U
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
The Product uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Product is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should bet hat of a typical
)
commercial or hospital environment.
T
If the user of the Product requires continued
operation during power mains interruptions, it
)
is recommended that the Product be powered
T
from an uninterruptible power supply or a
battery.
)
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.