3. General data
The accuracy relates to 1 year at a temperature between 18 °C – 28° and 75 %
humidity (yearly calibrations are offered).
Autoranging and manual measurement range selection
Max. voltage between the connection socket and ground: 300 V DC/AC
Fusing:
Max. operating height:
Height of display:
Display:
Polarity indicator
Overrange indicator:
Sampling rate:
Low battery status:
Automatic power off
Power supply:
Operating temperature:
Storage temperature:
Dimensions:
Weight:
Test Standard:
Measurement category
Definition of measurement categories.
Measurement category II: Measurements on circuits directly connected to low volt-
age networks electrically via plug. Typical short-circuit current < 10 kA.
Measurement category III: Measurements within the building installation (stationary
consumer devices with non-plug-in connection, distributor connection, permanently
installed equipment in the distributor). Typical short-circuit current < 50 kA.
Measurement category IV: Measurements at the source of the low voltage installa-
tion (meters, mains connection, primary overcurrent protection). Typical short-
circuit current >>50 kA.
To establish the measurement category in a combination of test lead and measur-
ing instrument, the lowest category, either of the test lead or the measuring instru-
ment, always
applies.
F 200 mA(4 A)/500 V quick-blow fuse
2000 m above MSL
200 mm, LCD
Max. 1999 (3½)
Automatic
'OL' is displayed
Approx. 0.4 s
Battery symbols is displayed
After approx. 15 min
2 x 1.5 V AAA Micro
0 °C to 40 °C
-10 °C to 50 °C
143 x 72 x 33 mm
400 g incl. batteries
IEC/EN 61010-1
:
CAT III 300V
- 17 -