Télécharger Imprimer la page

Knick PolyTrans P32000P0/11-KTA Mode D'emploi page 53

Convertisseurs universels

Publicité

Les langues disponibles

Les langues disponibles

Configuration using Switches: Example
Sensor:
Range:
Output signal:
1. Adjust sensor type:
TC Type J:
2. Adjust start value:
200 °C
This start value is composed of: numerical value = 20, factor = x10.
Adjust numerical value using rotary switches:
Adjust factor x10:
3. Adjust end value:
1200 °C
For end values above 1000 °C, adjust factor x10+1000
Adjust numerical value using rotary switches:
Adjust factor x10+1000:
4. Adjust output signal:
4 ... 20 mA:
Thermocouple type J
200 ... 1200 °C
4 - 20 mA
DIP1 = 1, DIP2 = 0, DIP 3 = 0
DIP7 = 0, DIP8 = 1
Please note:
You find an adjustment aid at
www.knick.de/Polytrans:
"Intuitive Adjustment"
20
DIP4 = 0, DIP5 = 1
20
DIP6 = 1
Caution!
After completion of con-
figuration you must cover the
switches with the included
self-adhesive polyimide tape.
53

Publicité

loading

Ce manuel est également adapté pour:

Polytrans p32000