Installation
Your satisfaction and safety require a complete understanding of this unit. Read the
instructions thoroughly and be sure all operators are given adequate training before
attempting to put the unit in use.
NOTE: This equipment must be used only for its intended application; any alterations or modifications
will void your warranty and may cause injury.
Intended use
VWR INCU-Line Tower is a general purpose convection incubator with two separate chambers, capable of
uniformly cooling and heating samples from 10 ºC below ambient up to a maximum of 50 ºC. The unit has two
transparent doors, so samples can be observed without opening the doors, thus preventing temperature
changes in the chambers.
For research use only. Not intended for any animal or human therapeutic or diagnostic use.
Symbols and conventions
The following chart is an illustrated glossary of the symbols that are used in this manual.
CAUTION
This symbol indicates a potential risk and alerts you to proceed with caution.
Product Specifications
Specifications
Electrical
Maximum total power consumption
Power consumption per chamber at 15 ºC
Power consumption per chamber at 37 ºC
Power consumption per chamber at 50 ºC
Fuses
Controls
Temperature range
Temperature resolution (display)
Temperature accuracy at 37 ºC
Temperature fluctuation (time) at 37 ºC
Temperature variation (spatial) at 37 ºC
Environment temperature
Interior volume
Maximum load per shelf
Maximum load per chamber
Exterior dimensions W x D x H
Interior dimensions W x D x H
Weight
5
vwr.com I EN Instruction manual VWR
®
INCU-Line Tower
INCU-Line Tower
230 V, 50/60 Hz, 0.7 A
160 W
65 W
35 W
60 W
2 x 2 AT 250 V
PID control
Ambient -10 ºC to +50 ºC
0.1 ºC
± 0.5 ºC
± 0.2 ºC
± 1.0 ºC
+15 ºC to +40 ºC
2 x 23 l
3 kg
9 kg
380 x 470 x 790 mm
294 x 300 x 270 mm (each chamber)
34 kg