Maintenance
6.3.3 Replacing the respiratory filter
WARNING
No protection without respiratory filter!
► Do not use the device without respiratory filter.
CAUTION
Damage to blower unit due to penetration of particles!
► Make sure when you remove the respiratory filter that no
particles enter the device through the suction inlet.
The respiratory filter changing process may differ
depending on the respiratory filter type used.
Particle filter
Removing the respiratory filter:
1. Push respiratory filter lock button.
2. Fold out splash guard cover with respiratory filter (Figure
A).
3. Remove used respiratory filter (Figure B).
Inserting the respiratory filter:
1. Check rubber seal on respiratory filter for damage.
2. Insert new respiratory filter into splash guard cover so that
the respiratory filter is firmly seated in the splash guard
cover.
3. Hook the two lower connection points of the splash guard
cover at an angle into the blower unit sockets.
4. Fold splash guard cover in until it audibly snaps into place
at the respiratory filter lock button.
12
A
Gas cartridge or combination filter
Removing the respiratory filter:
1. Push respiratory filter lock button.
2. Fold respiratory filter with splash guard cover out (Figure
A).
3. Dismantling the splash guard cover:
a. Press on the centre of the upper splash guard cover
edge until it snaps out.
b. Fold out splash guard cover (Figure B).
Inserting the respiratory filter:
1. Check rubber seal on respiratory filter for damage.
B
2. Place splash guard cover with its lower edge at an angle
on the respiratory filter.
3. Push on splash guard cover until it snaps audibly into
place.
4. Hook the two lower connection points of the respiratory
filter at an angle into the blower unit sockets.
5. Fold respiratory filter in until it audibly snaps into place at
the respiratory filter lock button.
6.3.4 Checking flow and warning devices
1. Make sure that a respiratory filter is inserted.
2. Connect the plug-in connector of the breathing hose to the
blower unit.
3. Switch on the blower unit by pushing the
control panel for approx. 2 seconds.
 After it is switched on, the device performs a self-test.
If the device does not work properly or warning
devices are triggered, eliminate the fault.
4. Cover the open end of the breathing hose with the palm of
your hand.
 The blower unit starts operating more intensively after
about 5 seconds. An alarm is triggered after about
20 seconds.
Have the blower unit checked if the blower speed
remains unchanged and no alarm is triggered.
5. If you wish, you can switch off the blower unit by pushing
the
button on the control panel once again.
6.3.5 Replacing the O-ring at plug-in or bayonet-
type hose connector
1. Use the O-ring removal tool to lift the old O-ring out of the
groove.
2. Insert new O-ring in the provided groove.
3. If required, lubricate the new O-ring with Molykote 111.
Instructions for use
A
button on the
®
|
Dräger X-plore
B
8000