G. DIsInFectInG your PArI
reusABLe neBuLIZer
1. clean the nebulizer as described in section F . For low-
level disinfection every other treatment day, soak all
parts of the paRi Reusable Nebulizer (except tubing)
in a solution of 1 part distilled white vinegar and 3
parts hot tap water for 1 hour. You can substitute
respiratory equipment disinfectant (like control iii
for distilled white vinegar (follow manufacturer's
instructions for mixing).
2. Rinse all parts of the paRi Reusable Nebulizer
thoroughly with warm tap water and dry with a clean,
lint-free cloth. discard the vinegar solution when
disinfection is complete.
3. the paRi Reusable Nebulizer parts (except tubing)
may also be disinfected by boiling them in a clean pot
of water for a full 10 minutes.
4. hospitals and clinics can sterilize the paRi Reusable
Nebulizer parts (except tubing) by autoclaving (a
maximum of 277° F or 136° c), follow autoclave
manufacturer's instructions.
5. dry parts on a clean, lint-free cloth.
CONTROL III
is a registered trademark of Maril Products, Inc.
®
REMEMBER
Replace YouR Reusable NebulizeR cup eveRY siX MoNths
Advantages of Replacing Your Nebulizer Cup:
* Lester, Mary K. nebulizer use and Maintenance by cystic fibrosis Patients: a Survey Study. respiratory care 2004; 49(12): 1504-1508.
K
All manuals and user guides at all-guides.com
H. mAIntenAnce oF your
FILter
Your trek
use, the filter will slowly turn grey, indicating the need for
replacement. the filter must be replaced immediately if it
gets wet or becomes clogged. a good rule to follow is to
®
)
replace your filter every 6 months – at the same time as
your paRi reusable nebulizer .
Removing: First make sure the trek
turned off. using a coin, remove the air filter cover by
turning it counter-clockwise. pull the cap and filter out,
then remove the filter from the cap (Fig. j-K). take a new
filter and press into cap with the open end visible (Fig K-l).
place the cap and filter back into the front of your trek
use the coin to tighten by turning the cover clockwise.
note: Do not attempt to wash and reuse air filters. Do not
use substitute filter material; compressor damage may result.
comPressor & recHArGeABLe BAttery
use a clean, damp cloth to wipe the exterior of the
compressor housing. be careful not to get liquids onto the
power Receptacle or the battery contact since damage
could occur.
mAIntenAnce & rePAIr
Never operate your paRi trek
incompletely assembled or damaged. use only original
paRi parts and attachments authorized by paRi for the
paRi trek
parts and attachments may damage the unit or cause
improper treatment. do not open or disassemble the
compressor unit at any time.
note: there are no serviceable parts inside the Pari trek
compact compressor System. Service by anyone voids warranty.
• Consistent medication delivery to the lungs
• Fast and efficient treatments
• Bacterial contamination can be reduced*
L
Questions? 1.800.FAst.neB (327.8632) | PArI.com/treks
trek
s neBuLIZer system
®
®
s air filter is bright white when new. with
s if it is improperly or
®
®
s compact compressor system. use of other
®
s power switch is
s.
®
S
®
EN-8

Publicité

loading