Storage/Transport Environment
Temperature:
-40°C to +65°C (-220°F to 149°F)
Relative Humidity:
20% to 80% (non-condensing)
Altitude:
-15m to 10,600m
Pressure:
241hPa to 1060 hPa

Maintenance

The maintenance that can be performed by the
operator is limited to cleaning and disinfecting
the device. Any maintenance inside the device
must be performed by qualified service
personnel only.
• Before cleaning the device(s), switch off at
the mains.
• Clean the device with a clean, slightly damp
cloth, and wipe it dry.
CAUTION Do not use solvent silicon-based or
abrasive cleaning agents.
CAUTION Make sure that no liquids enter the
device through the connectors and other
openings in the enclosure.
WEEE – Waste Management
The device and its accessories must be disposed
of separately as electronic waste in accordance
with your local authorities.

Weight and Dimensions

• Weight: 10.9kg
• Dimension: 81x250x275mm (HxWxD)

Safety Checks

The following safety checks should be conducted
by qualified personnel at least once a year and in
the event of repair:
• Inspection for visible damage to device.
• Inspection of mains cord and connecting
cables.
• Possible: Measurement of insulation
resistance.
• Measurement of leakage currents.
• Measurement of resistance of protective earth
conductor.

Classification

Type of protection against electric shock:
• Class I: Equipment in which protection
against electric shock does not rely on basic
insulation only, but which includes an
additional safety precaution in that means are
provided for the connection of the equipment
to the protective earth conductor in the fixed
wiring of the installation in such a way that
accessible metal parts cannot become live in
the event of a failure of the basic insulation.
Method(s) of sterilization:
• Device not intended to be sterilized.
Degree of patient vicinity:
• Device suitable for use in patient
environment.
Degree of protection against electric shock:
• Enclosure protectively earthed (treated as
applied part Type B).
Degree of protection against harmful ingress of
water:
• IP20: Ordinary equipment (enclosed
equipment without protection against ingress
of water).
Degree of safety of application in the presence of
a flammable anesthetic mixture with air or with
oxygen or nitrous oxide:
• Equipment not suitable for use in the presence
of such a mixture.
Mode of operation:
• Continuous operation at normal load and
Momentary operation at high load.
• Momentary operation at high load is used for
short time printing with high power laser
printers.
• Momentary operation at high load is possible
for up to 60 sec per hour at elevated
temperatures.