Accessories cannot be repaired. Take care of them
Do not position legs, armbands, abdominal band or
other accessories close to direct or indirect sources
of heat.
Do not position the device in damp or wet areas. Do
not wet the device under any circumstances. Do not
pour any liquid on the device.
Do not clean the device or accessories with water or
detergents. Simply wipe with a damp cloth.
Do not store the device with its accessories in
particularly cold areas or in areas with temperature
below zero. It can be damaged seriously.
If you do not use the device for a long period, store
it in its box.
Usage environment and case
• Use and keep the device and its accessories in a clean environment with a temperature
between 10 – 40°C (50 – 104°F), moisture 30 – 85%.
• Do not use outdoors.
Problem found
The device does not
power up
A strange hiss similar
to that of an air leak is
perceived
The air does not arrive
at the applicators, there
is not enough pressure
The air continues to
enter the chambers and
an air leak in applicators
is perceived
• Check correct connection to electrical outlet.
• Make sure you have turned the switch (off-on (0 - I) button) to position
ON (I).
• Check the status of the power cord which must be intact.
• Check that all pipes, accessories and connectors are intact.
• Make sure that all connections (to the machine and the applicators) are
correctly inserted.
• Make sure that the air connection pipes are not bent, pinched or cru-
shed.
• Check the proper connection of connectors to the machine and the ap-
plicators.
• Make sure that the air connection pipes are not bent, pinched or cru-
shed.
• Make sure that the inflation of the chambers occurs progressively and
according to the chosen program.
• Make sure that you have set the pressure control correctly.
• Make sure there is actual air outlet from the machine.
• Check the presence of holes or damage in the air chambers of the ap-
plicators. If no anomalies are found, switch the device off, restart it and
check correct functioning again.