15. Do not use lubricants on this device or any of its accessories to avoid the risk of fi re
and burns.
16. Use of this device at an altitude above 4,000 m (13,000 feet), or outside the tem-
perature range of 5°C (41°F) to 40°C (104°F), or outside the humidity range of 5% to
93% may adversely a ect the fl owrate and percentage of oxygen and consequently
the quality of therapy. When not in use, the device should be stored in a clean, dry
environment between -20°C and 60°C (-4°F and 140°F). Use and/or storage outside
of the valid conditions may damage the product. For more technical details see
Chapter 15 - Technical Description. Allow this device up to 20 minutes to warm/cool
from storage temperatures to operating temperature before use.
17. Always ensure at least one battery is inserted before using this device.
18. If feeling ill or experiencing discomfort while using this device, contact your clinician
or seek medical assistance immediately to avoid harm.
19. Your home oxygen provider must verify the compatibility of the device and all acces-
sories used prior to use. To ensure you are receiving the therapeutic amount of oxy-
gen for your medical condition, the device and accessories must only be used after
one of more settings have been determined or prescribed for you at your specifi c
activity levels by a healthcare professional.
20. The electrical cord and tubing could present a tripping or strangulation hazard. Keep
away from children and pets.
21. Do not disassemble or modify this device or any of its accessories. Do not attempt
any maintenance other than tasks described in Chapter 10 - Troubleshooting. Disas-
sembly can create an electric shock hazard and will void the warranty. Contact your
distributor for servicing by authorised personnel.
22. Use only spare parts recommended by the manufacturer to ensure proper function
and to avoid the risk of fi re and burns.
23. Do not repair or perform service work while the device is in use by the patient.
24. Do not position your device and its power supply such that it is, di cult to unplug
from the power source in the event of an emergency or it is di cult to isolate the
device in the event of a fi re, electrical hazard or any other incident.