11. MAINTENANCE AND CLEANING
11.1. ROUTINE MAINTENANCE
WARNING: Do not use lubricants on this device or any of its accessories.
CAUTION: Replace the cannula on a regular basis. Check with your distributor or clini-
cian to determine how often the cannula should be replaced.
Device will indicate with an alarm when service is required. (Also, see Chapter 10 - Trou-
bleshooting.)
EN
No special maintenance needs to be carried out by the user. Contact your provider or
manufacturer for help, in setting up, maintenance, or to report unexpected errors.
11.2. CLEANING AND DISINFECTION
WARNING: Do not submerge this device in liquid. Do not expose to water or precipita-
tion. Do not expose to dusty conditions.
CAUTION: Do not use cleaning agents other than those specifi ed in this manual. Allow
the cleaning solution to dry from the cleaned surface before use.
CAUTION: Always disconnect power and turn o this device before cleaning.
Clean the exterior with a soft cloth slightly dampened with soapy water or with anti-bac-
terial wipes (Isopropyl alcohol 70% solution).
For disinfecting, use a MadaCide-FDW-Plus wipe or equivalent and follow the manufac-
turer's instructions. (Manufacturer - Mada Medical Products Inc., www.madamedical.com)
Important: The device should receive an external cleaning weekly, accessories should
be cleaned as needed. The device exterior and accessories should be cleaned and dis-
infected and the patient fi lter replaced prior to delivering to a new patient. The device
may become hot after operation, take additional care when replacing fi lters.
Nasal cannula: Refer to the original manufacturer's instructions for cleaning the nasal
cannula.
11.3. SERVICE LIFE
The expected service life of the device is 5 years, except for the sieve modules. The
service life of the sieve modules will depend on the operating conditions. The sieve mod-
ules are an internal component of the device and should only be replaced by a trained
person. If the intake and exhaust vents are not blocked and the check vents alarm is
activated contact your distributor.