Table 6-4
List of calibration items (continued)
Function
AC mV
[d]
Resistance
Diode
Keysight U1253B User's and Service Guide
Range
Calibration item
50 mV
3.000 mV (1 kHz)
30.000 mV (1 kHz)
30.000 mV (10 kHz)
500 mV
30.00 mV (1 kHz)
300.00 mV (1 kHz)
300.00 mV (10 kHz)
1000 mV
300.0 mV (1 kHz)
1000.0 mV (1 kHz)
1000.0 mV (10 kHz)
(done)
Short
SHORT
50 M
OPEN
10.000 M
5 M
3.0000 M
500 k
300.00 k
50 k
30.000 k
5 k
3.0000 k
500 
300.00 
(done)
Short
SHORT
2 V
2.0000 V (done)
Performance Tests and Calibration
[a]
Reference input
3 mV, 1 kHz
30 mV, 1 kHz
30 mV, 10 kHz
30 mV, 1 kHz
300 mV, 1 kHz
300 mV, 10 kHz
300 mV, 1 kHz
1000 mV, 1 kHz
1000 mV, 10 kHz
Dual banana shorting plug with
copper wire
Unplug all test leads or shorting plug,
and leave the terminals open
10 M
3 M
300 k
30 k
3 k
300 
Dual banana shorting plug with
copper wire
2 V
6