6
Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available,
substitute with another calibration standard of equivalent accuracy.
A suggested alternative method would be to use the Keysight 3458A 8½ – Digit
Digital Multimeter to measure less accurate yet stable sources. The output value
measured from the source can be entered into the instrument as the target
calibration value.
Table 6-1
Recommended test equipment
Application
DC voltage
DC current
Resistance
AC voltage
AC current
Frequency
Capacitance
Duty cycle
Nanosiemens
Diode
Frequency counter
Temperature
Square wave
Short
168
Recommended equipment
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Keysight 33250A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Keysight 33250A
Fluke 5520A
TM Electronics KMPC1MP
(K-Type thermocouple extension)
Keysight 53131A and Keysight 34401A
Pomona MDP-S
Recommended accuracy requirements
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
< 20% of U1253B accuracy spec
Keysight U1253B User's and Service Guide