Performance Tests and Calibration
Table 6-2
Performance verification tests (continued)
Step
Test function
7
Press
to select AC mV measurement
8
Turn the rotary switch to the
9
Press
to select conductance (nS)
measurement
10
Turn the rotary switch to the
176
Range
50 mV
[a]
500 mV
1000 mV
500 
position
[n]
5 k
50 k
500 k
5 M
50 M
500 M
500 nS
Diode
position
Reference signals/
values
50 mV, 1 kHz
50 mV, 10 kHz
45 mV, 20 kHz
45 mV, 30 kHz
45 mV, 100 kHz
500 mV, 45 Hz
500 mV, 1 kHz
500 mV, 10 kHz
450 mV, 20 kHz
450 mV, 30 kHz
450 mV, 100 kHz
1000 mV, 1 kHz
1000 mV, 10 kHz
1000 mV, 20 kHz
1000 mV, 30 kHz
1000 mv, 100 kHz
500 
[n]
5 k
50 k
[n]
500 k
5 M
50 M
[d]
450 M
50 nS
[e]
1 V
Keysight U1253B User's and Service Guide
Error limits
± 0.24 mV
± 0.39 mV
± 1.695 mV
± 1.695 mV
± 1.695 mV
± 2.25 mV
± 2.25 mV
± 2.25 mV
± 16.95 mV
± 16.95 mV
± 16.95 mV
± 6.5 mV
± 11.5 mV
± 47 mV
± 47 mV
± 47 mV
[c]
± 350 m
± 3 
± 30 
± 300 
± 8 k
± 505 k
± 36.10 M
± 0.6 nS
± 1 mV