1. Press SET.
2. Select [SYSTEM] and press CAL.
3. Select [DATE/TIME] and press CAL.
4. Select [MODIFY] and press CAL.
5. Adjust date and time and press CAL.
1. Press SET.
2. Select [SYSTEM] and press CAL.
3. Select [PASSWORD] and press CAL.
4. A private code can be programmed to avoid undesired access to the instru-
ment. Choose [YES] and press CAL to enter your secret sequence of 5 keys.
1. Press SET.
2. Select [DATA] and press CAL.
3. Select [INTERVAL] and press CAL.
4. Select the desired time interval between the data-logging and press CAL.
View the stored values:
1. Select [TABLE] and press CAL.
2. Select the desired channel by pressing CAL.
3. Select the first line to view and press CAL.
4. Select [SHOW TABLE] and press CAL.
Send the stored values:
1. Select [SEND] and press CAL to send the stored data to the digital port.
Erase the stored values:
1. Select [ERASE] and press CAL. Follow the instructions on the screen.
1. Press SET.
2. Select [4-20 mA] and press CAL.
3. Select the desired range and press CAL.
4. Select the desired minimum level corresponding to a 4 mA recorder output and
press CAL.
5. Select the desired maximum level corresponding to a 20 mA recorder output
and press CAL.
1. Press SET.
2. Select [RS485] and press CAL.
3. Select [INTERVAL] and press CAL.
4. Select the desired time interval between the data transmission and press
CAL.
5. Select [ID#] and press CAL.
6. Select the desired identification number (0...999) and press CAL. The instru-
ment can be identified by e.g. a computer when a specific number is allocated
to it.
7. Select [BAUD] and press CAL.
8. Select the desired baudrate and press CAL.
R361 • R362 • R363

Date and time

Password

Data-logging

(R362, R363)

Analog output

(R362, R363)

RS485

(R362, R363)
4