DA-GEN DRYDEN AQUA GENERATOR Mode D'emploi page 5

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
10.1
pH module
Check if the chip is plugged in
correctly and if the green LED
is blinking.
(PH/RX label on the left side)
Temperature module
Temperature
probe
1
red
yellow
2
3
black
Free Chlorine control
1
2
3
4
Free Chlorine probe
red
black
3
4
Chlorine probe detector
black
3
FL2
(rotameter)
brown
5
blue
6
If using a Variable Speed
Pump, calibrate the probe
using the lowest filtration
speed.
Check if the chip is
plugged in correctly
and if the green LED
is blinking
(CL label facing down)
10. MEASUREMENTS
10.2
10.1 pH Calibration
10.11
pH
10.2 Temperature calibration
10.21
1 2 3
10.3 Free Chlorine calibration
10.31
1
2
3
4
5
6
7
8
10.36
4
10.3
10.1 Measurements: Adjustment of setpoints and measuring probes.
10.2 Setpoints for each measurement.
10.3 Setpoints:
pH: 7.4-7.6; redox: 600 - 800 mV; free chlorine: 0.1- 0.5 ppm; conductivity:
~ 2000 μS/cm
10.12
10.13
10.14
10.15
10.22
10.22 Temperature calibration: To set the difference
between the measured value of the probe and the actual
temperature, use the up/down keys. Set to the actual
temperature of the probe and press OK.
Let the system run 24 hours before calibrating and make sure
to have a free chlorine level of at least 0.3 ppm free chlorine!
10.32
10.33
10.31 Calibration of the Free Chlorine probe:
Recommended at least every 2 to 3 months
10.32 Calibration with buffer (photometer
DPD1): Follow the instructions in 7 steps that
appear in the display.
10.33 Step 1 of 7 - Calibrate Cl at 0 ppm
(offset): Close the water flow through the probe
and wait for 5 to 60 min until the reading is close
10.34
10.35
to 0. Press OK
10.34 Step 3 of 7 - Calibrate Cl: Set the water
flow to the correct rate of 80-100 litres/hour.
Wait for 1 to 10 min until there is a stable ppm
reading. Press OK.
10.35 Step 5 of 7 - Establish the real ppm
values with the up/down keys according to your
DPD1 (free chlorine) value. Press OK.
10.36 Step 7 of 7 - If this screen is not shown
10.37
10.38
repeat the calibration process.
10.38 Manual calibration: Open the water flow
and set the flowmeter (rotameter) to the correct
flowrate (50-100l/h). Wait until the current level
is stable. Set the chlorine level with the up/
down keys, manually (use a manual DPD1 test
kit). Press OK when value is correct.
10.11 Calibration of pH probe: Recom-
mended at least every 2-3 months during the
usage season. Calibrate it always first with
the Buffer (2pt).
10.12 Calibration with buffers (buffer pH7 /
pH10 / neutral): Follow the instructions that
appear on the display.
10.14
Manual
calibration:
manual adjustment of the probes – only
recommended to correct small deviations in
the readings.
10.15 Without removing the probe from the
water, use the up/down arrows to adjust the
reading so it matches your reference value
(photometer or other measurement).
Allows
EN