8 LOCKING AND UNLOCKING THE CALIBRATION (continued)
To unlock the calibration:
1
Press the Cal softkey and select Calibration Lock.
2
Enter the four digit PIN Code, if set, using the
select 0 to 9 and the  softkey to move from the first to the fourth
digit .
3
Press Ok or Escape to cancel.
To disable the calibration lock PIN Code:
1
Press the Menu softkey and select Setup/Calibration Lock.
2
Enter the four digit PIN Code.
Should the user forget or loose the PIN code, the PIN can be disabled via
®
ElcoMaster . Using the USB cable, simply connect the gauge to a PC with
®
ElcoMaster Version 2.0.33 or higher installed and select Edit/Clear Calibration
PIN.

9 CALIBRATION METHODS

CALIBRATION
METHOD
Zero
Smooth
Rough/ 2 Point
Zero Offset
The  softkey will appear when the " X " changes to a number.
www.elcometer.com
MODEL
DESCRIPTION
B, S, T
A one point calibration - ideal for smooth
surfaces, simply place the probe on the
uncoated substrate and the gauge will
automatically calibrate
B, S, T
A two point calibration - the user selects a
suitable calibration foil and a smooth,
uncoated substrate (zero)
B, S, T
The ideal calibration method for rough or
profiled substrates, using two foils of known
thickness - one above, another below the
target dry film thickness
S, T
Calibration method for measurement when
the substrate profile/ roughness is unknown,
or not accessible. A user defined offset value
is applied to the measurement reading
softkeys to
R