6 TAKING A READING (continued)
DO NOT
Drag the probe over the coated surface.
Bang the probe down hard onto the surface.
Allow the probe to hover over the surface as this could result in
a false reading.
The display will dim if inactive for more than 15 seconds and will go 'black' if
inactive for the period defined in Menu/Setup/Screen Settings/Screen Timeout.
Either press any key, or tap the gauge to awaken it.
The gauge will switch off automatically after 5 minutes of inactivity.
- - - indicates reading outside range of probe.

7 CALIBRATING THE GAUGE

1
Press the Cal softkey.
2
For alternative calibration methods,
select Cal/Cal Method.
3
Select Calibrate and follow the on-
screen instructions.
4
When prompted place the probe on
the centre of the foil(s).
Foils can be stacked.
Not all calibration methods are available for all gauge types see 9 Calibration
Methods.
8 LOCKING AND UNLOCKING THE CALIBRATION
To set a calibration lock PIN code:
1
Press the Menu softkey and select
Setup/Calibration Lock.
2
Set the four digit PIN Code using the
softkeys to select 0 to 9
and the  softkey to move from the
first to the fourth digit .
3
Press Ok, Escape to cancel or Adjust
to amend the PIN code.
The  softkey will appear when the " X " changes to a number.