4. Turn On the Multimeter

Press the Power (on/off) switch:
The front-panel display illuminates while the
multimeter performs its power-on self-test.
Following self-test, the multimeter defaults to
DC voltage measurements.* Typical display:
Measurements are shown in the primary
display (bottom left) and the measurement
range is shown in the secondary display (top
right).
*You can set the multimeter to return to the last
power-down state when power is applied. See the
34405A User's and Service Guide for details.
6. Set the Range
For most measurement functions, you can let
the multimeter automatically select the range
using autorange or you can select a fixed
range using manual ranging. The ManRng
annunciator is displayed when manual
ranging.
Selects a higher range and disables
autoranging
Range
Selects a lower range and disables
autoranging
To return to autoranging and disable manual
ranging, press:
Shift
34405A Quick Start Guide
mV DC
Range
mV DC
5. Select a Function
Press one of these keys to select a
measurement function:
DCV
ACV
DCI
ACI
The continuity and diode test functions are
shifted (shown in blue). For example, to
select continuity, press:
Shift
DCV
Press
to return to the DC voltage
function.
7. Set the Resolution
You can select either 4½ or 5½ digit
resolution for most measurement functions.
5½ digit readings have the best accuracy and
noise rejection. 4½ digit readings provide for
faster readings.
4
Digits
Shift
Selects 4½ digit mode
Digits 5
Shift
Temp
Selects 5½ digit mode
Note: The continuity and diode test functions
have a fixed, 4½ digit display. Capacitance
and temperature have a fixed, 3½ digit
display.
Cont
)
)
)
Freq
Freq
Temp
Cont
)
)
)
5