All manuals and user guides at all-guides.com
12
4.2.2 Connecting the VP cable to the transmitter
Standard connection for CO
transmitter 5100 e
2
Function
VP cable wire
Terminal connection
pH glass
transparent
1
Reference
red
2
Temperature 1
white
E
Temperature 2
green
D
Shielding
green/yellow
C
For further details check instruction manual of
the transmitter.
4.2.3 Calibation of pH electrode / transmitter 5100 e
Min. immersion depth
25 mm / 1"
Prepare electrode as described in section 4.1 "Start-
up". We recommend a 2-point calibration in pH buffer
pH 7.00 and pH 9.21. Before calibration, first remove
wetting cap. Normally you start with buffer pH 7.00 for
determining zero point and then buffer pH 9.21
for determining the slope. Please refer to instruction
manual of the transmitter for further details.
Important settings: Temperature sensor select Pt1000.
InPro 5000
52 002 465
InPro 5000 CO
Sensor 12 mm
2
CO
Transmitter
2
VP Cable
© 09 / 06 Mettler-Toledo AG
Printed in Switzerland