c) Commissioning and use
Connecting the Suprasson handpiece.
See the instructions supplied with the handpiece.
Mounting the tips
To obtain adequate vibration of a tip it must be
correctly tightened and not forced beyond its
stop. To avoid it becoming locked, the tip should
be removed after each use.
See the Tipbook for the power settings for
SATELEC tips.
Connecting the micro-motor
To connect the micro-motor to its cord connector,
simply screw it on.
The two red marks should line up.
Where a prophylactic contra-angle handpiece is
being used, close the spray tap on the front of the
unit to stop the irrigation flow.
The speed of the micro-meter is adjusted using
the power button (Fig 1-3) on the front, from 1 to
10.
There is a three-way switch on the face of the
unit (Fig. 1-1) to change the direction of the
motor:
In the up position: rotation is in a clockwise
direction.
In the down position: rotation is in an anti-
clockwise direction.
In the center position: the motor is stationary.
Note: Do not change the direction of rotation
with the footswitch depressed. This will cause
the motor to stop. You must wait 1 minute for
the unit to reset.
IMLN motor
Only an approved Satalec technician may adjust
the light intensity of the IMLN motor. The time lag
for the light is 3s after release of the footswitch.
Use of the 3-function syringe
The syringe may be used in conjunction with
other functions and it works independently of the
footswitch.
There are 3 possible combinations:
air alone
water alone
air+water (atomised spray)
d) WARNING
It is essential to turn off the water supply
before leaving the dental surgery.
8