Malfunc-
Possible reasons
tions
Err -1
Movement of the arm or body during the meas-
urement. Do not talk during the measurement.
Err -2
Pump problem
Err - 3
Deflation rate too high or too low, valve problem
Err - 300
Upper pressure limit of 300 mmHg has been
exceeded.
Irregular heartbeat, disruptive movements,
shaking, wobbling (objects), breathing in deeply,
etc.
Levels too
Was the necessary rest observed before
high
measurement?
Unusual
Movement or talking during measurement. Rest-
levels
ing time not observed, feet possibly crossed,
refrain from smoking or drinking coffee at least
30 minutes before measurement.
All manuals and user guides at all-guides.com

Corrective action

Repeat the measurement after 3-5 minutes. Don't move
the arm and don't talk. In case of recurrence please
contact our after-sales service.
Check whether the air connection is correctly inserted. Is
the cuff fitted too loosely? Repeat measurement. In case
of recurrence please contact our after-sales service.
In case of recurrence please contact our after-sales service.
Repeat the measurement after 3-5 minutes rest
Repeat measurement after 3-5 minutes rest. Possible
factors (depending on the severity of the arrhythmia)
affecting the measurement results should be discussed
with your doctor.
Repeat measurement after a break of approx. 3-5 min.
Do not move your arm and do not talk.
Check conditions and repeat measurement.