Specifications
Model:
Operation principle:
Display:
Measuring range:
Blood pressure
measurement:
Laboratory accuracy:
Clinical accuracy:
Inflation:
Deflation:
Power supply:
Operating temperature:
Storage temperature:
Humidity:
Cuff:
Clinical validation:
If device is not used within specified temperature and humidity ranges the technical
accuracy can not be ensured.
Type BF equipment
Subject to change without notice.
This device conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
22
BP 1750 / BP 1700
Oscillometric method
Liquid Crystal Display
0–320 mmHg (cuff pressure)
40–199 / min (pulse rate)
20 mmHg (minimum diastolic value)
300 mmHg (maximum systolic value)
± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
According to AAMI-SP10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Automatic
Step valve
2 batteries, type LR 03 (AAA)
+ 10 °C to + 40 °C
– 20 °C to + 60 °C
Up to 85% RH
Suitable for wrist circumference of 13 to 21 cm
According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to different results for devices
which are calibrated against auscultatory reference.)