Electrostatic
discharge
(ESD) IEC
61000-4-2
Power
frequency
(50Hz)
magnetic
field
IEC 61000-
4-8
Guidance and manufacture's declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration-electromagnetic immunity
The CMS50D Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of CMS50D Pulse
Oximeter should assure that it is used in such an environment.
Immunity
IEC60601
test
test level
Radiated
3V/m
RF
80MHz
ICE
to
61000-4-
2.5GHz
3
±6KV contact
±8KV air
3A/m
Compliance
level
3V/m
±6KV
contact
±8KV air
3A/m
Electromagnetic
environment -guidance
Portable and mobile RF
communication equipment
should be used no closer to any
part of the CMS50D Pulse
Oximeter, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
recommended separation
distance
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment