Chlorine With Liquid Reagent - Tintometer AL250 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Methods
GB
0.0.0
Zero
Test
CL
RESULT
!
Zero
Test
CL
RESULT
30

Chlorine with liquid reagent

0.01 – 4.0 mg/l
a) free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Remove the vial from the sample chamber and empty the vial.
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
Add the water sample to the 10 ml mark. Close the vial tightly with
the cap and invert several times to mix the contents.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Immediately after measurement add
3 drops of DPD 3 solution
to the already coloured test solution. Close the vial tightly with the
cap and invert several times to mix the contents.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 27)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
> 2 – 3 mg/l: ± 0.20 mg/l
> 3 – 4 mg/l: ± 0.30 mg/l
Notes:
see page 28
AL250_Cl-Br-pH-CyA-TA-CaH_2a 05/2011
marks
marks

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Lovibond checkitdirectAqualytic al250

Table des Matières