Unscrew the container while holding the cap in place; empty the used oil into a suitable
container for used oils. Screw the container back on while holding the cap in place. Carefully
replace the container in its lodging without exerting pressure on the scale. N.B.: In order to
avoid damage to the oil scale, never exert pressure on it from above or below.

REPLACING THE PRINTER PAPER

Use only heat-sensitive paper of the type described below.
Paper width: 2.28inchs (58 mm)
Maximum paper roll diameter: 1.26 inches (32mm)

CHECKING SCALE CALIBRATION

Turn unit on and note the "Gas Available" reading. Hang provided 1 lb weight from the hook
found on the bottom of the scale (see fig. 4.) If change in "Gas Available" reading is +/- 1 oz, it
is recommended that the scale be calibrated.

UNSCHEDULED MAINTENANCE / CALIBRATION

Perform this operation whenever the values displayed on the LCD do not correspond to real
values. WARNING: The operations listed below must be performed with maximum attention
and care. In particular, always observe the following precautions. Always place the weights at
the center of the scale plate. Never exert pressure on the oil scale. Always recover the gas in
the high- and low-pressure hoses before beginning calibration of the transducers.
To enter the CALIBRATION MENU:
Switch the machine on. The MAIN MENU will appear after a few seconds:
Select NEXT MENU:
Now select CALIBRATION from the menu:
Type the code 0791 and press ENTER to confirm.

CALIBRATING THE BOTTLE SCALE

Disconnect the machine from the electrical supply. Procure a known reference weight from
55.6 lb to 71.1 lb (28 to 32 kg). Remove the plastic cover on the rear of the machine to access
the machine bottle. Close the blue and red taps on the bottle. Unscrew the bottle lock nut (3 –
Fig. 4). Separate the heating coil (r – Fig. 8) from the bottle (do not touch or disconnect the
wires of the resistance coil). Remove the bottle (d – Fig. 8) from its lodging, leaving the
resistance around the scale plate. Set the bottle on a stand at least 15 inches (40 cm) in
height.
INSTRUCTIONS MANUAL
AUTOMATIC PROCEDURE
ASSISTED PROCEDURE
NEXT MENU
Gas Avail
x:xx lb
<CALIBRATION>
DATA AND CONFIGURAT.
SERVICE ALARM
BOTTLE FILLING
To enter calibration
insert code
....
And press ENTER