Commands
NAMUR
Function
Commands
IN_PV_1
Read the external actual temperature
IN_PV_2
Read the internal actual temperature
IN_PV_3
Read the safety actual temperature
IN_PV_4
Read the pump actual speed
IN_SP_1
Read the internal setting temperature (if 0: internal control)
Read the external setting temperature (if 1: external control)
IN_SP_3
Read the safety setting temperature
IN_SP_4
Read the pump setting speed
IN_TMODE
Read temperature control
0: internal control
1: external control
OUT_SP_1 xxx
Set the internal setting temperature XXX (if 0: internal control)
Set the external setting temperature XXX (if 1: external control)
OUT_SP_12@n
Set the WD safety temperature with echo of the set (defined) value.
OUT_SP_4 xxx
Set the pump speed XXX
OUT_SP_42@n
Set the WD-safety speed with echo of the set (defined) value.
IN_STATUS
Status bits
OUT_TMODE_0
Set to Internal temperature control
OUT_TMODE_1
Set to External temperature control
OUT_WD1@n
Start the watchdog mode 1 and set the time for the watchdog to n (20...1500) seconds.
Echo of the Watchdog time.
During a WD1-event, the heating and pump functions are switched off.
This command needs to be send within the watchdog time.
OUT_WD2@n
Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump set
speed is set to the WD safety speed.
This command needs to be send within the watchdog time.
RESET
Reset the PC control and stop the device functions.
START_1
Start the heating function and stop the device functions.
START_4
Start the pump function
STOP_1
Stop the heating function
STOP_4
Stop the pump function
PC 1.1 Cable
This cable is required to connect RS 232 port (10) to a PC.
1
6
2
7
3
8
4
9
5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
Fig. 26
9
5
9
4
8
3
7
2
6
1
1
2 RxD
3 TxD
PC
4
5 GND
6
7 RTS
8 CTS
9
46