IKA IC control Mode D'emploi page 44

Table des Matières

Publicité

Note: When the "PROGRAMS" is started, the M1 seg-
ment settings have higher priority.
Default Status:
Defines the default state (OFF) of the external valve as
"Open" or "Closed".
Refill:
Uses the external M1 valve to automatically check the
level when operating with water.
Observe "Default Status" setting.
Low level: Sets the valve switch-on point (ON).
High level: Sets the valve switch-off point (OFF).
Note: Prior to using the "Refill" option check the func-
tion of the buoyage (4).
The "Refill" option has higher priority than the program
M1 segment settings.
M2-Output:
Alarm:
Activates the alarm output (switch contact).
Error: In the event of an error the "M2-Output —
Switch" is activated (ON, inverted initial state).
Warning: In the event of a warning the "M2-Output
— Switch" is activated (ON, inverted initial state)
When "Warning" is activated, "Error" is simultane-
ously activated.
Default Status: Defines the default state (OFF) of the
"M2-Output — Alarm" as "Open" or "Closed".
Switch:
Activates the switch output.
ON: Switches output into the active state (inverted ini-
tial state).
Note: When the "Program" is started, the M1 seg-
ment settings have higher priority.
Default Status: Defines the initial state (OFF) of the
switch output as "Open" or "Closed".
M3-Input:
Ext. Standby:
External standby input to stop the device functions Tem-
per and Pump.
Activated: Activates the ext. standby function. Device
functions are stopped in the event of an inverted initial
state (ON) at the input.
Default Status: Defines the default state (OFF) of the
input as "Open" (high level) or "Closed" (low level).
OUTGAS
In the option, the user is allowed to set the outgas speed
in range of 1000 rpm to 4600 rpm and outgas interval
from 10 seconds to 240 seconds.
This function can be used when filling external devices
such as laboratory reactors.
SAFETY
Time out:
In the menu "Time out", the user can determine a
time limit in the event of a communication breakdown
between the station and the Wireless Controller. The
station continues to work with the preset target values
until the preset time value has expired. Following that,
the station runs using the preset safety temperature and
safety speed.
Note: The initial time out is 30 seconds and the user
can define up to 60 minutes for this time limit.
Speed:
In the "Speed" menu, the user can specify a appropri-
ate and safe speed for specified temperature.
Note: The factory setting of the safe speed is 1000 rpm
and is activated after the time limit is set (see "Time
out").
Temperature:
In the "Temperature" menu, the user can specify a
temperature that is appropriate and safe for the circu-
lating task.
Note: The initial safe temperature is 30 ºC and is acti-
vated after the time limit is set (see "Time out").
Password:
In the "Password" menu, the user can protect the
Wireless Controller settings using a 3 digits password.
When accessing the menu screen, the user is prompted
for a password. 000 deactivates the password prompt
(factory setting).
SETTING
Languages:
The "Language" option allows the user to select the
desired language.
Display:
The "Display" option allows the user to change the
background color and brightness of the working screen.
Sound:
The "Sound" option allows the user to activate/deacti-
vate the key tone.
44

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hbc 5 controlHbc 10 control

Table des Matières