Wavetek Meterman 85XT Manuel D'utilisation page 11

Table des Matières

Publicité

DC AND AC CURRENT MEASUREMENT (See Fig. 2)
❶ Connect red test lead to the mA Input for current measurements up to 200mA
or to the 20A input for current measurements to 20A (15 to 20A for 30 seconds).
Connect black test lead to COM Input Connector. ❷ Set the Function/Range
switch the desired DC or AC current range corresponding to the selected input
jack. If in doubt choose to the highest setting first. ❸ Open circuit in which
current is to be measured. Securely connect test leads in series with the load. ❹
Turn on power to circuit being tested. š Read current value on Digital Display.
Note: Each current input has a protective fuse installed. If fuse does blow, replace
with same.
Note for AC Measurements: See AC/DC Voltage measurement.
Discharge capacitors
3b
Kondensatoren entladen
Descargue los condensadores
D charger les condensateurs
Fig. 2
3a
4
3d
A
5
DC
ON/OFF
V
1000
200
20
2
200m
2
200
2m
20m
20A
200m
A
20M
2M
200k
20k
2k 200
True RMS
V-
COM
CAT.
II
200mA MAX
FUSED
MAX
1000V
750V
- 9 -
A
AmA
HOLD
750 200
V
20
A
2
200m
200m
20m
20A
2m
200
200
k
20k
2k
DUTY%
Hz
85XT
mA
20A
MAX
20A/60sec
FUSED
1
red/rot/
roja/rouge

Publicité

Table des Matières
loading

Table des Matières