OHAUS Frontier FC5714 Manuel D'instructions page 45

Centrifugeuse
Masquer les pouces Voir aussi pour Frontier FC5714:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EN-38
11.6 Table 6: Error messages
1.800.561.8187
TM
Frontier
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R
Error-No.:
Description
1
Imbalance arose
2
Imbalance sensor is defective
4
Imbalance switch has been activated for longer than 5 seconds
8
Transponder in the rotor is defective
11
Temperature sensor is defective
12
Chamber over temperature
Leap of speed is too big between two mesaurements
14
CLOSE lid
33
Open lid while motor is running
34
Lid contact defective
38
Lid motor is blocked
40
Communication with frequency converter distrubed during start
41
Communication with frequency converter distrubed during stop
42
Short circuit in the frequency converter
43
Undervoltage frequency converter
44
Overvoltage frequency converter
45
Over temperature frequency converter
46
Over temperature motor
47
Over current frequency converter
48
Timeout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
70
Timeout between controler and RS232 interface
99
Rotor is not allowed in this centrifuge
Inserted rotor does not exist in the programm
FALSE
Rotor is not detected
rotor no