Télécharger Imprimer la page

Masimo NomoLine Luer Serie Mode D'emploi page 3

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
NomoLine® Luer Series
Single-Patient Use Sampling Lines
Single-Patient Use Only
INDICATIONS
The NomoLine® Product Family is indicated for the measurement of respiratory rate and respiratory and anesthetic gases in
adult, pediatric, infant and neonatal patients. The NomoLine Product Family includes single-patient use sampling lines for
gas sampling and/or oxygen delivery. The NomoLine Product Family is indicated for use by clinical professionals in healthcare
environments, including mobile environments.
WARNINGS
• Carefully route the NomoLine Sampling Line to reduce the risk of patient entanglement or strangulation.
• Do not apply negative pressure to remove condensed water.
• Dispose in accordance with local regulations for biohazardous waste.
• Do not submerge the NomoLine Sampling Line in any cleaning solution or attempt to sterilize by autoclave, irradiation,
steam, gas, ethylene oxide or any other method as this may seriously damage the NomoLine Sampling Line.
• Do not re-use due to cross contamination risk.
• Always replace the NomoLine Sampling Line if an occlusion alarm occurs.
• Do not use NomoLine Cannulas on patients with obstructions in one or both nostrils.
• The CO
measurement can be diminished if NomoLine Cannulas are used during mouth breathing or on a patient that
2
has a perforated septum.
• Do not use NomoLine Adult/Pediatric Airway Adapter Sets for infants/neonates as the adapter adds 6 ml dead space.
• Do not use NomoLine Infant/Neonatal Airway Adapter Sets for adults/pediatrics as this may cause excessive flow
resistance (0.7 ml dead space).
• Do not use NomoLine Airway Adapter Sets with inhalers or nebulized medications as this may clog the bacteria filter.
DESCRIPTION
NomoLine Sampling Lines with Luer include nasal and nasal/oral cannulas for non-intubated patients and airway adapter
sets for intubated patients. The NomoLine Sampling Lines with Luer are accessories to sidestream gas analyzers with a luer
connector. Refer to the compatible monitor's User's Guide for technical specification and full system use instructions.
INSTRUCTIONS
• Open the bag and remove the NomoLine with Luer Sampling Line. For convenient preparations, do not yet remove the
tape that keeps the tubing coiled.
A. Fitting a NomoLine Cannula with Luer to the patient
• If fitting a NomoLine Nasal CO
• If fitting a NomoLine Nasal/Oral CO
the oral scoop in front of the mouth.
• If fitting a NomoLine Single Nasal Prong CO
• Refer to Fig. 4: Pass the Cannula lines over the ears (1) and adjust the slider for comfort and fit under the chin (2). (This
step does not apply to NomoLine Single Nasal Prong CO
B. Connecting a NomoLine Airway Adapter Set with Luer to the patient circuit
• Refer to Fig. 5: Connect the small diameter end of the Airway Adapter to the Y-piece of the ventilation source.
• Refer to Fig. 6: Connect the large diameter end of the Airway Adapter to the patient's endotracheal tube.
C. Connecting a NomoLine Sampling Line with Luer to the gas analyzer
• Refer to Fig. 7: Remove the tape that keeps the tubing coiled. Turn the NomoLine with Luer sampling line connector
into the device's gas sampling inlet until it can't be turned anymore.
• Ensure that the tubing is properly connected and that it is not twisted or crimped.
D. Connecting a NomoLine Cannula with Luer to an oxygen source
• Refer to Fig. 8: Remove the tape that keeps the tubing coiled. Attach the oxygen supply tubing to the oxygen source.
Adjust the oxygen flow. (This step applies only to NomoLine with Luer Cannulas with O
• Ensure that the tubing is properly connected and that it is not twisted or crimped.
REPLACEMENT
NomoLine Sampling Lines with Luer are single patient use products and should after use be disposed of in accordance with
local regulations for biohazardous waste and replaced for each new patient. They should also be replaced if the sampling line
becomes occluded.
The lifetime of the product depends on the application and product type.
DI R E C T I O N S F O R U SE
LATEX
Not made with natural rubber latex
PCX-2108A
02/13
Cannula with luer refer to Fig. 1: Insert the prongs into the nostrils.
2
Cannula with luer refer to Fig. 2: Insert the prongs into the nostrils and position
2
2
Cannula with luer refer to Fig. 3: Insert the prong in one of the nostrils.
Cannula with luer.)
2
3
.)
2
10102B-eIFU-0920
en

Publicité

loading