The ARTROMOT®-SP can be used
on the left and on the right ankle joint.
However, the device must be converted
first. This is done very quickly.
1.
Activate the "transport setting" func-
tion
in the menu (see also 5.3)
and start the ARTROMOT®-SP .
2. Remove the footplate (7) as well as its
bracket (11). To do so, pull it from the
square-end guide at motor B. (Fig. G)
3. Loosen thumbscrew (12) used to mount
motor B to the moving bracket (15).
Then pull the motor upward and re-
move (Fig. H).
4. Mount motor B on the opposite side of
the carriage, using the thumb- screw.
5. Push the footplate bracket into the
square-end aperture on motor B.
Observe the marks: red dot on square
rod of the footplate bracket to match
red dot on square-end aperture on
motor B, mark L or R on the footplate
bracket to match mark L or R on the
motor housing.
Note!
– To assist in the conversion, motor B
is equipped with a side identification
function.
The following symbols may appear
on the display during conversion: (ex-
ample) conversion from left to right:
motor on the left-hand side:
motor removed from bracket:
conversion to the right-hand side
completed:
– If, after the conversion, the footplate
bracket and thus the footplate is at
an inadmissible angle, the display
shows:
Please press the START key and the
carriage will automatically move to
the permitted range and stop.
. Environmental Protection Statement
The product described in this operation
manual must not be disposed of with un-
sorted household or municipal waste.
It requires separate disposal. Please con-
tact DJO Global for information about the
possible recycling of the product.
. Specifications
ARTROMOT®-SP
Model:
ARTROMOT®-SP Standard, Part no.: 80.00.035
Device name:
ARTROMOT®-SP Standard Chip, Part no.: 80.00.036
ARTROMOT®-SP Comfort, Part no.: 80.00.037
ARTROMOT®-SP Comfort Chip, Part no.: 80.00.038
Input ratings:
100 – 240 V
tolerance -15% to +10%
Current consumption:
100V
standby (ON):
5VA
operation (maximum):
33VA
330mA
Fuses:
2x T A H250Vac interrupting rating 1500A
according to IEC 60127-2/3, UL 248-14, CSA
C22.2 no. 248.14
serial numbers <20,000: I
Protection class:
serial numbers >20,000: II
serial numbers <20,000: IPX
IP class:
serial numbers >20,000: IP
type B
Applied part:
20 kg
Max. continuous load on carriage:
Dimensions (transport):
length:
78.0 cm
width:
42.0 cm
height:
39.5 cm
The service life of the product as well as
of the supplied parts and accessories is 6
years minimum.
/ 50 – 60 Hz
240V
5VA
33VA
140mA