DJO Global ARTROMOT-SP3 Mode D'emploi page 34

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
Note!
– If a pause is programmed when you
enable the function, the carriage will
stop for the duration of the pause
each time the maximum stretching
value is attained.
– Enabling the stretching in inversion
function automatically disables mo-
tor A. The only movement performed
is inversion / eversion.
– It is not possible to combine a second
stretching function with the special
function „stretching in inversion".
– During operation the display will
show the symbol
.
stretching in eversion
With the special "stretching in eversion"
function the carriage gently stretches
the joint towards the outside.
Starting at the middle position the car-
riage will first move to the programmed
inversion value and then to the pro-
grammed eversion value.
Subsequently the carriage reverses 5°
towards inversion and then moves very
slowly back again to the programmed
eversion value (display:
for the right
foot and
or the left foot). After that
it attempts to attain another 5°, moving
even slower than before (display
or
the right foot and
for the left foot).
If a high resistance toward the ad-
ditional 5° is sensed, the reverse on
load function is automatically activated
and the carriage moves in the opposite
direction.
This stretch cycle is repeated ten times.
After that the carriage moves to the
programmed maximum inversion value
and restarts the eversion stretch cycle.
default setting: disabled
Note!
– If a pause is programmed when you
enable the function, the carriage will
stop for the duration of the pause
each time the maximum stretching
value is attained.
– Enabling the stretching in eversion
function automatically disables mo-
tor A. The only movement performed
is inversion / eversion.
– It is not possible to combine a second
stretching function with the special
function „stretching in eversion".
– During operation the display will
show the symbol
.
Reminder:
You save the selected parameter values by
pressing the STOP key.
. Care, Maintenance, Transport, Conversion
. Care
Warning!
Shock hazard –
Remove the power cord from the wall
outlet before cleaning.
Shock hazard, equipment damage –
Liquids must not enter the device or the
programming unit..
– ARTROMOT®-SP can be disinfected
by wiping down with a disinfectant.
Thus, it complies with the special
hygiene standards for medical technical
equipment.
– The frame and the removable leg
supports can be cleaned with com-
mon disinfectants and mildhousehold
detergents.
– Only use a damp cloth to wipe the CPM
device down.
Warning!
Patient hazard –
patient contamination
– Before using the device on another
patient, be sure to clean and disinfect
it according to the instructions given
here.
Caution!
Equipment damage –
– The plastic material used is not
resistant to mineral acids, formic acid,
phenols, cresols, oxidants and strong
organic or inorganic acids with a pH
value below 4.
− Use only clear disinfectants to
prevent discoloration of the device.
− Do not expose the CPM device
to strong ultraviolet radiation
(sunlight) and fire.
. Maintenance
(fuse replacement)
Check before each use
Visually inspect the device for signs of
mechanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.