– pauses:
0
– timer:
continuous
operation
– speed:
50 %
– reverse on load - motor A: 25
– reverse on load - motor B: 25
– motor A:
ON
– motor B:
ON
– synchronized mode:
enabled
– total therapy time:
0
– special functions:
disabled
LEVEL :
reverse on load - motor A
(safety circuit)
In the synchronized mode, the carriage
automatically moves to the middle posi-
tion when the patient's resistance (load)
exceeds the set value.
In the non-synchronized mode, the
carriage automatically moves in the
opposite direction when the patient's
resistance (load) exceeds the set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25 at level 1, very little resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
default setting: level 25
reverse on load - motor B
(safety circuit)
In the synchronized mode, the
carriage automatically moves to the
middle position when the patient's re-
sistance (load) exceeds the set value.
In the non-synchronized mode, the
carriage automatically moves in the
opposite direction when the patient's
resistance (load) exceeds the set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25 at level 1, very little resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
default setting: level 25
Caution!
Patient hazard –
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked joints
and similar situations. The manufac-
turer cannot be held liable for misuse of
this feature.
motor A ON/OFF
To allow an isolated movement, the mo-
tors can individually be turned on and
o . Motor A performs dorsal extension
and plantar flexion movement, motor B
performs eversion and inversion.
For an isolated eversion / inversion
movement, set motor A to the appropri-
ate position (dorsal extension / plantar
flexion), then turn the motor o .
In the normal operating mode, the dis-
play will show the symbol for the dorsal/
plantar movement
disabled
.
default setting: motor A ON
motor B ON/OFF
To allow an isolated movement, the mo-
tors can individually be turned on and
o . Motor A performs dorsal extension
and plantar flexion movement, motor B
performs eversion and inversion.
For an isolated dorsal extension /
plantar flexion movement, set motor B
to the appropriate position (inversion /
eversion), then turn the motor o .
In the normal operating mode, the
display will show the symbol for the
inversion / eversion movement
disabled
.
default setting: motor B ON
Note!
Please note that one motor (A or B)
must always be turned on. Otherwise
the display will show the message:
LEVEL :
transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select the
function and press the START key. The
carriage moves to the transport posi-
tion. In the top left-hand corner of the
display you will see
(also refer to
chapter 6 Transport).
synchronized/non-synchronized mode
Motors A and B can run in synchronized
or non-synchronized mode.
synchronized:
Motors A and B perform a synchronized
movement according to the ankle joint's
physiological motion pattern:
From the middle position the carriage
first goes to the maximum eversion
value. Afterwards, the carriage will
simultaneously move to the maximum
inversion value and to the maximum
plantar flexion value, and then simulta-
neously to the maximum dorsal exten-
sion value and to the value halfway
between inversion and eversion. After
reaching this position, the cycle restarts:
maximum eversion at the same time
as halfway between dorsal extension /
plantar flexion, etc.
During operation the synchronized
mode is indicated in the top left-hand
corner by the symbol
.
non-synchronized:
Both motors run independently of each
other within the selected ranges of
motion.
To enable the non-synchronized mode,
disable the synchronized mode.
During operation the non-synchronized
mode is indicated in the top left-hand
corner by the symbol
.
default setting: synchronized mode
enabled
Caution!
Patient hazard –
The synchronized operation is
recommended as a general rule.
However, medical and/or therapeutic
indications may call for the non-
synchronized mode. In the non-syn-
chronized mode, particular caution and
attention on the part of the physician/
therapist is required to rule out any risk
for the patient.
total therapy time
ARTROMOT®-SP without patient
chip card
The total therapy time is the added sum
of operating hours.
If the device is used by only one patient,
this time is equivalent to the duration of
all the patient's therapy sessions.
Under menu item „total therapy time"
of ARTROMOT®-SP models with
chip card you can view each patient's
total therapy time (duration of all the
patient's therapy sessions).
Deleting the stored therapy time
Press and hold the parameter key for
5 seconds or select the New Patient
function.
Service MENU
For service purposes only, refer to
Service Manual.
Reminder:
You save the set parameter values by
pressing the STOP key.