DJO Global ARTROMOT-SP3 Mode D'emploi page 32

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
. Programming
ARTROMOT®-SP Comfort
ARTROMOT®-SP Comfort devices o er
further functions at two additional pro-
gramming levels.
The programming levels are selected in the
same way as on Standard models.
Programming levels 1, 2, 3 and 6 are identi-
cal with programming levels 1 to 4 of the
Standard models.
With the default settingsall special func-
tions are disabled.
The following treatment values, settings
and information can be entered/viewed on
the programming unit (22):
LEVEL :
– dorsal extension
– plantar flexion
– inversion
– eversion
MENU
LEVEL :
– pause
– therapy timer
– speed
– new patient
MENU
LEVEL :
– reverse on load -
motor A
– reverse on load -
motor B
– motor A ON/OFF
– motor B ON/OFF
MENU
LEVEL :
– warm-up protocol
– isolation program
– therapy documentation
dorsal extension /
plantar flexion
– therapy documentation
inversion / eversion
MENU
LEVEL :
– stretching in dorsal
extension
– stretching in plantar
flexion
– stretching in inversion
– stretching in eversion
MENU
LEVEL :
– transport setting
– synchronized/
non-synchronized mode
– total therapy time
– service menu
. Protocol details –
Comfort models
– To access the programming level you
need, press the MENU key repeatedly.
– The treatment parameters are selected
with the di erent parameter keys.
– To change the treatment values, press
the + / – keys.
– To enable/disable a function, press the
respective parameter key again.
– To save the entered values, press the
STOP key.
With the default settings all special
functions are disabled.
Note!
LEVEL : equivalent to level 1
of the Standard model (see: 5.3)
LEVEL : equivalent to level 2
of the Standard model (see: 5.3)
LEVEL : equivalent to level 3
of the Standard model (see: 5.3)
LEVEL : equivalent to level 4
of the Standard model (see: 5.3)
LEVEL :
warm-up protocol
During warm-up, the patient will slowly
become used to the set maximum dorsal
extension / plantar flexion and inversion
/ eversion values, starting from the
middle position.
The warm-up protocol starts in the
middle between the two maximum
values set for dorsal extension / plantar
flexion and inversion / eversion.
In the beginning, the range of inversion
/ eversion motion increases with each
cycle, until the programmed maximum
values are reached after a total of 15
cycles. In this phase, motor A (dorsal
extension / plantar flexion) is inactive.
Then the carriage stops again in the
middle position and starts increasing
the range of dorsal extension / plantar
flexion motion with each cycle, until
the programmed maximum values are
reached after a total of 15 cycles. In this
phase, motor B (inversion / eversion) is
inactive.
After warm-up the device enters the
normal mode.
During warm-up, the display will show
the symbol
in the upper, left-
hand corner.
default setting: disabled
isolation program
For this special function, both motors
are turned on, but they will never move
simultaneously.
These are the steps of the special
function:
– At first motor B moves to the
programmed maximum values for
inversion and eversion for ten cycles,
then it stops. In this phase, motor A is
disabled (symbol for motor A:
)
– Then motor A will move to the pro-
grammed maximum values for dorsal
extension and plantar flexion for ten
cycles and stops. The full cycle will
start over with motor B. Motor B is
disabled while motor A is operating
(symbol for motor B:
)
– The stop position at the end of each
tenth cycle (for both dorsal exten-
sion / plantar flexion and inversion
/ eversion) can be selected in steps
of 25% between 0% and 100% of the
programmed maximum range of
motion.
– Steps 1 and 2 can be repeated as
often as needed. You can stop treat-
ment with the STOP key; after the
programmed therapy duration, it will
stop automatically
During warm-up, the display will show
the symbol
in the upper, left-hand
corner.
default setting: disabled
Follow these steps to program the spe-
cial function:
– Access menu level 4, then press the
parameter key to select the special
function
.
– The display will show the symbol
for the special function, its status
(circle with or without check mark) as