Threaded
The Probe is available in three thread types:
2" NPT, 2" BSP or PF2.
Note:
Before inserting The Probe into its
mounting hole, ensure that the threads are of
the same type to avoid damaging The Probe
threads.
Flange Adapter (optional)
The Probe can be fitted with the optional 75 mm
(3") flange adapter for mating to 3" ANSI, DIN
65PN10 and JIS 10K3B flanges.
Sanitary
Notes:
• Mount The Probe so that the face of
the sensor is at least 25 cm above
the highest anticipated level.
• The Sanitary Probe 3A approval is
for chemical clean-in-place
applications to 60 °C (140 °F) only.
Ensure your cleaning chemicals are
compatible with ETFE or PVDF,
depending on your Probe type
• mount The Probe onto the top of
the tank's sanitary ferrule
• secure mating by surrounding
the joint with the clamp
• tighten adjusting wing nut
Note:
Inside of sanitary
ferrule must be smooth, free of
burrs, seams or ridges.
Page 2
*supplied with the Probe
The Probe – INSTRUCTION MANUAL
hinged lid
7 mm
(0.3")
117 mm
(4.6F")
cable entry
'knock out'
279 mm
(11")
87 mm
(3.4") ø
mounting
thread
61 mm
(2.4")
optional
The Probe
flange
adapter
7 mm
hinged lid
(0.3")
117 mm
(4.6")
cable entry
'knock out'
279 mm
(11")
87 mm
(3.4") ø
76 mm
integral
sealing ring
61 mm
(2.4")
Sanitary Ferrule,4" FDA Approved
The Probe
(US Food and Drug Administration)
clamp*
adjusting
wing nut
ferrule
tank
dimensions are approximate
84 mm
(3.3") max.
electronics
mounting
thread
74 mm
sensor
(2.9")
53 mm
(2. 1 ")
ø max.
84 mm
(3.3") max.
electronics
mounting
thread
sensor
(2.9")
53 mm
(2. 1 ")
ø max.
119 mm (4.68")
97 mm (3.83")
7ML19981GD61