LEM ISO plus Manuel D'utilisation page 14

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Basic operation

Connect the test leads to the test object and check for external voltages.
Select the resistance measuring mode by turning the rotary switch to the Ω
position.
If the external voltage is greater than 3V a warning symbol is displayed. Even
if a live voltage is present, the fuse will not be damaged.

Test lead compensation can be activated by connecting the leads together
and pressing the
key. The instrument is now in a Relative measurement
mode indicated by the ZERO announciator on the display. This mode is
also valid for continuity testing. The actual value of compensation R
indicated on the small display.
Press the
key to activate the continuity test with a 200mA test current,
and an audible alarm buzzer when measuring less than the adjusted
threshold.
ƒ
Adjust continuity level using the
selects the digit,
increments the digit
Pressing the
key again exists the LIMIT setting mode and activated the
new limit.
zero
key from 0,2 to 50 Ohm,
- 14 -
is

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Si1300z

Table des Matières