IV. 6. Technical characteristics
Dimensions: L x W x H: 368 x 400 x 724 mm.
Weight: 24.5 kg.
Noise level: ±48 dBA
Flow values:
0 - 5 l/min.
Accuracy of flow supplied:
In compliance with the ISO 8359 standard, the flow supplied is equal to the flow
set on the flow selector, accurate to within ± 10 % or 200 ml/min., whichever is
the larger of the two.
Average oxygen content:
• at 2 l/min. 93 %.
• at 4 l/min. 91 %.
• at 5 l/min. 90%.
(values at 21 °C and at an atmospheric pressure of 1013 mbar). Max. recom-
mended flow: 5 l/min. The variation of the maximum recommended flow does
not exceed ± 10 % of the indicated value when a counter pressure of 7 kPa is
applied to the output of the device. The maximum outlet pressure is 630 kPa.
Electrical power supply:
• 230 V - 50 Hz Europe / 115 V – 60 Hz / 230 V – 60 Hz (other countries,
depending on version)
• Use the cable provided, and check that the electrical characteristics of the
mains socket used match those indicated on the manufacturer's plate on the
rear panel of the machine.
• Mean power rating: 350 watts
Class II
Type B
Protected by a 2.5 A circuit breaker switch (230 V).
Protected by an 8 A circuit breaker switch (115 V).
Filters:
At the rear of the device: a dust filter.
Before the oxygen outlet: a product filter < 2.0 µm.

Air circulation:

A fan cools the compressor compartment.
Environmental limit conditions:
The performances of the device (especially the oxygen concentration) are
quoted at 21 °C and 1013 mbar. They may change with temperature and
altitude.