Introduction And General Description; Suction Lift; Setup - Masterflex E/S 07571-00 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

®
The MASTERFLEX
E/S™ (Environmental Sampling) Pump Drive provides pump speeds from 70 to 400 rpm in a clock-
wise or counterclockwise direction for purging the tubing and for the sampling of liquids.
The Drive and charging system are current limited, protected from reverse power supply polarity and transient voltages.
You may operate the Drive on the internal battery, an external 12V DC source or 115V AC or 230V AC when used with the
AC/DC Power Supply/Converter supplied with the unit.
The Drive has a built-in spill-proof battery and recharging system allowing the unit to be recharged from an AC power
source using the supplied AC/DC Power Supply/Converter or from an automotive 12V DC electrical system if used with
the Automotive Power Adapter accessory.
It is designed to accept and power one MASTERFLEX
Pump Head for fluid transfer and sampling in the field.
The unit is housed in a high visibility, protective housing. The control panel and the drive controls are protection rated to
IP54 only when a Pump Head is mounted on the drive.
When closed, the drive will float for a minimum of 30 minutes if dropped into a lake or stream to allow recovery of the unit.
The combination of the MASTERFLEX
ple of water up to twenty-six (26) feet vertically. This capability is called suction lift and is measured from the surface of the
liquid to the Pump Head.
If the liquid to be sampled has a viscosity or density greater than that of water, this lifting capability is reduced or dimin-
ished.
1. Care has been taken in the packaging of the drive to protect it in transit from the factory to its final destination. After
opening the carton, save the packing material until proper product operation has been verified.
The drive consists of the following items:
(1) Drive Unit
(1) AC/DC Power Supply/Converter
(2) Keys for the locking latches
(1) Operating Manual
(1) Warranty Card
(1) Pouch for Accessories
CAUTION:
Use of pump heads, tubing sizes and formulations other than those specified in this manual, or the
mounting and use of two or more pump heads concurrently will result in improper operation and
possible damage to the unit.
WARNING:
Turn drive off, remove all the power to the unit, including the AC/DC Power Supply/Converter, Au-
tomotive Power Adapter or other external power source if present before removing or installing
pump head or tubing. Place the Mode of Operation Switch (RUN/CHARGE) in the "CHARGE" posi-
tion and all other switches in the off position. This will help prevent accidental activation of the
drive mechanism so fingers or loose clothing will not get caught in the pump drive mechanism.
2. The drive does not include a Pump Head or tubing. The drive is capable of running one (1) of the following Pump
Heads: MASTERFLEX
See Appendix A for Pump Head and tubing catalog numbers recommended.
See Appendix B for flow rates and lift.

INTRODUCTION AND GENERAL DESCRIPTION

SUCTION LIFT

®
E/S™ drive and a single MASTERFLEX Pump Head is capable of lifting a sam-
®
®
L/S
Standard, L/S QUICK LOAD, L/S EASY-LOAD or L/S PTFE Tubing Pump Head.
®
®
L/S
Standard, QUICK LOAD

SETUP

4
®
®
, EASY-LOAD
or PTFE Tubing

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

E/s 07571-05

Table des Matières