1. The Drive will not op er ate on the in ter nal bat tery power source.
A. Ver ify that the Power Source Se lec tion Switch (B), is in the BAT TERY po si tion.
B. Ver ify that the Mode of Op er a tion Se lec tion Switch (C), is in the RUN po si tion.
C. Ver ify that the pump head, tub ing size and for mu la tion are cor rect and rec om mended for this drive.
D. If us ing a PTFE pump head, in stall and start the tub ing per the in struc tions in the pump head man ual. Ex ces sive
tight en ing of the oc clu sion bed can jam the drive mech a nism. Tighten the oc clu sion bed grad u ally as in di cated in
the pump head man ual.
E. Ver ify that the Pump Ro ta tional Se lec tion On/Off Switch (H), is ei ther in the CW or CCW po si tion for the
desired ro ta tion. The Drive Power In di ca tor (K) in the Drive Con trols sec tion will be il lu mi nated.
F. Tog gle or ac ti vate the Bat tery Sta tus In di ca tor (F). If the in di ca tor is RED, re charge the bat tery im me di ately.
You have drained or ex hausted the available power of the in ter nal battery.
G. If the Bat tery Sta tus In di ca tor (F) does not il lu mi nate, re move the con trol panel and check for loose con nec tions
to the bat tery.
H. Use a dig i tal multi-meter (DMM) set to DC volts to check the bat tery po ten tial. If be low 9 volts DC, re place bat tery.
I. If none of the above so lu tions work, contact your au tho rized dealer for ser vice.
2. The Drive will not run us ing ex ter nal power sup plied by the AC/DC Power Sup ply/Con verter.
A. Ver ify that the Power Source Se lec tion Switch (B), is in the EX TER NAL po si tion.
B. Ver ify that the Mode of Op er a tion Se lec tion Switch (C), is in the RUN po si tion.
C. Ver ify that the pump head, tub ing size and for mu la tion are cor rect and rec om mended for this drive.
D. If us ing a PTFE pump head, in stall and start the tub ing per the in struc tions in the pump head man ual. Ex ces sive
tight en ing of the oc clu sion bed can jam the drive mech a nism. Tighten the oc clu sion bed grad u ally as in di cated in
the pump head man ual.
E. Ver ify that the Pump Ro ta tional Se lec tion On/Off Switch (H), is ei ther in the CW or CCW po si tion for the de -
sired ro ta tion. The Drive Power In di ca tor (K) in the Drive Con trols sec tion will be il lu mi nated.
F. Ver ify that the co ax ial plug is fully in serted in the Ex ter nal Power In put of the unit.
G. Ver ify that AC volt age is pres ent in the AC out let to which the AC/DC Power Sup ply/Con verter is con nected.
H. Ver ify the DC volt age at the 2.5 mm co ax ial con nec tor. Must be at least 13 volts DC. AC/DC Power Sup ply/Con -
verter is not a ser vice able unit and con tains no user ser vice able parts.
I. If none of the above so lu tions work, con tact your au tho rized dealer for ser vice.
3. The Drive will not op er ate us ing the Au to mo tive Power Adapter Ac ces sory.
A. Ver ify that the Power Source Se lec tion Switch (B), is in the EX TER NAL po si tion.
B. Ver ify that the Mode of Op er a tion Se lec tion Switch (C), is in the RUN po si tion.
C. Ver ify that the pump head, tub ing size and for mu la tion are cor rect and rec om mended for this drive.
D. If us ing a PTFE pump head, in stall and start the tub ing per the in struc tions in the pump head man ual. Ex ces sive
tight en ing of the oc clu sion bed can jam the drive mech a nism. Tighten the oc clu sion bed grad u ally as in di cated in
the pump head man ual.
E. Ver ify that the Pump Ro ta tional Se lec tion On/Off Switch (H), is ei ther in the CW or CCW po si tion for the de -
sired ro ta tion. The Drive Power In di ca tor (K) in the Drive Con trols sec tion will be il lu mi nated.
F. Ver ify that the co ax ial plug of the Au to mo tive Power Adapter is fully in serted in the Ex ter nal Power In put of the
drive.
G. Ver ify that the Au to mo tive Power Adapter is fully in serted into the cig a rette lighter on the vehicle.
H. Ver ify that the fuse for the cig a rette lighter is not open or blown.
I. Ver ify the DC volt age at the 2.5 mm co ax ial con nec tor. It must be at least 12 volts DC.