„Manuel Mode" menu
Note: If the timer has been configured, then the time will start to
count backwards. If a timer has not been configured then the time
elapsed since the start is shown.
48
Note: Distillation sequences saved in this manner can be repeated
at a later dater in the user-defined mode by selecting the corre-
sponding procedure.
Note: If it is necessary to work in manual mode for an extended pe-
riod (>4 minutes) without performing a distillation (e.g. flask not in
heating bath), or if a drying process is carried out, then "Drying"
mode must be activated in menu "Settings", menu option "Distil-
lation", in order to prevent the "No temperature increase" error
message being displayed.