1. HEINE mini 3000
Batteries:
• 2 alkali cells (AA/ LR 6)
• The HEINE mini 3000
rechargeable battery 2,5V.
Charging time:
• The mini NT indicates when the rechargeable battery is fully-charged.
The illuminated ring is lit continuously.
• mini 2Z rechargeable battery (2,5V, NiMH) = ca. 4h (730mAh)
Instrument connector:
Internal thread in handle head (1).
Switching on:
Push switch (2) down.
Switching off:
Push switch up. If the handle is put back firmly into a coat pocket the handle is automatically switched off.
Changing the batteries:
Unscrew the end-cap (3) and shake out the old batteries / rechargeable battery. Insert new batteries /
rechargeable battery with the plus contact facing the handle head.
2. HEINE mini 3000
Changing the bulb:
Unscrew the instrument from the handle and pull the bulb (4) downwards. Wipe the new bulb glass clean and
insert it so that the locating pin (5) fits the slit.
3. HEINE mini 3000
Method of operation:
Attach the instrument to a handle. Switch on the illumination and direct the light emerging from the prism (7)
onto a white surface. Using the slide (1) you can adjust the focusing of the light from parallel (slide at bottom) to
convergent rays.
Use of magnifying lens +3D:
The lens is mounted on the cylindrical part (8) of the hand lamp. Ensure that the shoulder (5) of the lens ring (6)
is pointing upward. The lens enables the presbyopic examiner to carry out examinations in comfort and produces
a magnified fundus image. Always position the lens at the center of the hand lamp.
Changing bulbs:
Please note: The performance of this instrument can only be guaranteed if genuine HEINE bulbs are used.
Remove the instrument from the handle. Grip the collar (4) of the bulb between your thumb and index finger nails
and remove the bulb by pulling it downward. Wipe the dome of the new bulb on a clean cloth and then insert it so
that the adjusting pin (4) fits into the recess (2).
4. HEINE mini 3000
Fitting and ejecting the tongue-blade:
Push the HEINE disposable tongue-blade into the holder (3) until it clicks in. Press the trigger (4) to eject.
Changing the bulb:
Unscrew the cover (1) and pull out the bulb (2). Insert a new bulb and screw in the cover again.
5. HEINE mini 3000
HEINE mini 3000
Safety in use:
The otoscope should only be used to examine the ear when a tip is fitted. The contact time (skin) has to be limited
to max. 1 minute. Pressure should be applied with great care when doing pneumatic testing.
Attaching a tip:
The otoscope may only be inserted into the auditory canal when a disposable (grey) or reusable (black) tip is fitted.
Push the tip onto the instrument so that the internal projection engages in the slit (1). Twist to the right to lock.
10/32
Battery Handle
®
battery handle can be charged in the HEINE
®
Ophthalmoscope
®
Focalux hand lamp
®
tongue-blade holder
®
Otoscope
®
F.O. Otoskop
®
med 3206
mini NT with the HEINE mini 2Z
®
2013-12-05