20 Compliance Information for EMC Test

Emission test
Conducted and radiated
RF emissions
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Power
frequency (50/60Hz)
magnetic field
IEC61000-4-8
Electromagnetic immunity to RF wireless communications equipment
Test frequency
Band
(MHz)
385
380-390
450
430-470
710
745
704-787
780
Electromagnetic emission
Compliance
Group 11, Group 1,
Class B
Electromagnetic immunity
IEC 60601 test level
± 8kV contact
±15kV air
10 V/m
80 MHz to 2.7 GHz
30A/m
a
) (MHz)
Service
TETRA 400
GMRS 460,
FRS 460
LTE Band 13,17
Electromagnetic Environment – guidance
Home Healthcare environment
Compliance Level
± 8kV contact
±15kV air
10 V/m
80 MHz to 2.7 GHz
30A/m
a
b
)
Modulation
Pulse
b
modulation
)
18Hz
c
FM
)
±5kHz
deviation
1kHz sine
Pulse
b
modulation
)
217Hz
16
European union approval
Electromagnetic Environment –
Special healthcare environment
The relative humidity should be
at least 50%.
Home healthcare environment
Home healthcare environment
Maximum
Distance
)
power
(m)
(W)
1.8
0.3
2
0.3
0.2
0.3
guidance
Immunity test
level
(V/m)
27
28
9