Step 6- Select one of the seven stimulation
modes
Press the MODE button to toggle between
stimulation modes:
Mode1, Mode 2, Mode 3, Mode 4, Mode 5, Mode
6 and Mode7. LCD Screen indicates the current
mode.
Step 7- Select the stimulation times
Press the TIME button the stimulation The
stimulation time is 20 minutes each power on.
Timer range is 5 to 20 min.
Step 8- Adjust the stimulation intensity
Press the
button will Decrease & Increase the
intensity of stimulation.
Note: With the increase of intensity, you may
experience sensations like tingling, vibration, pain,
etc. Therefore, gradually increase the intensity,
and stop increasing when a comfortable level is
reached.
Step 9- Press the ON/OFF button once to turn
the power off
After the device is turned off. Disconnect the wires
and electrode pads. Take the batteries out of the
unit if the device will not be in use for a long period
of time.

13 Accessories

The following accessories are included:
Connecting Wires
The lead wires consist of PVC material in the outside and thin copper wire on the inside. The
total length of the lead wire is 59.05 inches and there are 3 connectors on each wire. The one
connector on the left end is used to connect to the device and the 2 connectors on the other
end are used to connect to the electrode pads. The pulse generated by the device to the pads
through these connecting wires.
Electrode Pads (Up to 80 uses)
The electrode pads applied to MDTS100 is adopted the electrode pad manufactured by Wuxi
Jiajian Medical Instrument Co., Ltd and has been cleared by FDA in June 23, 2009 as K090198.
12