Quick Start Guide CellaCast PA 8x
Two-colour channel [Qu] -> ATD function [Q . A td]
Parameter
EN
[T . D EL]
(T . A CT)
► Press [Mode] to access the main menu
> The display shows [Ao]
► Press the [▼] until [Qu] is displayed
► Press [Mode]
> The display shows [Q . E PS]
► Press [▼] until [Q . A TD] is displayed
► Press [Mode]
> The display shows [t. d El]
► Press [▼] until [t. A ct] is displayed
► Press [Mode]
> The current sampling time is displayed
► Press [▲ or ▼] for 2 second
> The display flashes three times
► Press [▲ or ▼] to set the desired sampling time
► Press [Mode] to confirm the parameter
Exit operation
► Wait 30 seconds
or
► Press [▲ or ▼] to change to the parameter [End]. Then Press [Mode] to change to the
functional menu.
► In the functional menu press [▲ or ▼] to change to the parameter [End], then press
[Mode].
Alternatively, the parameters can be set via IO-Link.
10 Adaptation of temperature level
By changing the emissivity ratio, the difference between the measured temperature level and
true temperature can be compensated.
For this purpose, a comparison measurement with an immersion lance is required. Adjust
emissivity ratio in such a way that the measured values coincide. An increase in the emissivity
ratio results in a low temperature value.
The graph shows an approximation of the influence of the emissivity ratio setting on the change
in temperature value.
Function
Time delay [s]
Sampling time [s]
Default
1 s
15 s