Charging the lift unit

RiseAtlas can be used with three different charging systems:
1. Charger station for charging via hand control (Standard version for 50100057)
2. In-rail charging (Accessory for 50100057)
3. End-point charging (Standard version for 50100051, accessory for 50100057)
RiseAtlas is charged with a charger unit that is connected to the charger cables via a DC outlet. The charger should be
kept in an appropriate location. Note that the charger is a separate unit and can be easily mistaken for another charger.
Ensure that the correct charger is used, see Technical information. For assembly instructions, see Assembly of charger unit.
1) The lift unit is charged via a wall-mounted charger station. Place the hand
control in the charger station. The charging lamp on the lift unit indicates charging
for 15 minutes after being activated. Charging continues until the battery is fully
charged.
2) For in-rail charging, the lift unit is always kept charged, which means that it can
be left anywhere on the rail and will charge automatically.
3) For end-point charging, the lift unit is placed at the charger station at the end
of the rail. If the rail system is of the traverse type, the secondary rail is placed at
the end of the primary rail at which the charger station is located. Check to ensure
that the battery symbol on the lift unit lights up.