Direct Healthcare Group SystemRoMedic RiseAtlas450M med QuickTrolleySystem Manuel Utilisateur page 7

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 164
Assembly of the trolley
Clearence hole for safety pin
To be able to mount and dismount the lift unit from the rail with the QuickTrolleySystem, the saftey pin, marked red in
Figure 1, needs a clearence hole in the rail that enables the saftey pin to slide up and down. Mounting and dismounting
the lift unit from the rail will only be possible at the position where the clearence hole has been made.
I. Decide where on the rail the lift unit can be mounted and dismounted. NOTE! Distance, D in Figure 2, from end of rail
to trolley depends on which end stop that is used, Table 1.
II. Drill a hole, diameter 10-12 mm, in the rail where the safety pin should slide up and down, Figure 3.
III. Remove metal chips from the rail.
VI. Put the accompanying label on the rail to show the user where the lift unit can be mounted and dismounted, Figure 4.
1; Safety pin
End stop used
Distance, D
Stop bult
50 mm
Adjustable end
120 mm
stop
Table 1; End stop
Safety pin
Place the safety pin in the intended position in between the
charger plates, Figure 1. The length of the safety pin chosen
depends on the height of the rail, Table 2.
SystemRoMedic
M A N U A L
TM
2; Trolley in position for mounting /
dismounting
4; Label on rail
Height of rail
H64
H120 – H160
Table 2; Length of safety pin
D
Safety pin
Short safety pin
M3x15 mm
Long safety pin
M3x25 mm
3; Trolley in rail
5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

5010005750100051Systemromedic riseatlas450m

Table des Matières