Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 48
Directions for Use

Cautions

A caution statement in this manual identifies a condition or practice, which if not
corrected or discontinued immediately, could lead to equipment failure, equipment
damage, or data loss.
WARNING For safety, the Video Colposcope should only be coupled to a
grounded 110–120 VAC hospital-grade outlet (220–240 volt, 50 cycle
international).
WARNING The lamp operates at a high temperature. DO NOT attempt to
remove the lamp before allowing it to cool. Allow at least five minutes for the
lamp to cool before replacing. Replace with Welch Allyn lamp #09800-U only.
Caution Federal law restricts sale of this device to, or to the order of, a
physician or other appropriately licensed medical professional.
Caution Occasionally inspect the power cord for signs of cuts, abrasions or
dents.
Caution The Video Colposcope should never be stored or operated in areas
where it could get wet or could be exposed to any environmental conditions like
extreme temperature or humidity, direct sunlight, dust, etc.
Caution All service to the Video Colposcope must be performed by Welch Allyn
or by an authorized repair center.
Caution There are no user servicable parts (other than the lamp and fuses) in
this unit or in its accessories. Any attempt to disassemble and/or repair this unit
will result in voiding of the warranty.
Caution The Video Colposcope is cooled via a fan located in the back of the
unit. The fan draws air in from beneath the Video Colposcope and exhausts the
air out the back of the Video Colposcope. To avoid overheating, verify that the
unit is no less than 6" from a wall.
Caution Do not clean illumination lens with alcohol. Do not touch optical or
illumination lenses except as described in Maintenance section of this manual.
Caution The colposcope can be damaged if the unit is transported while
holding the handle. The unit should be transported by grasping the pole.
Caution Do not sterilize.
Caution Do not spray or allow solution to drip into the air vents.
Caution Do not immerse any part of the unit in cleaning solutions.
Introduction
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

88001a89000a8800789001a88002a88004a ... Afficher tout

Table des Matières