Specifications

Guidance and Manufacturer's Declaration

Emissions and Immunity Information

The REF 880 and 890 Series Video Colposcope is intended for use in the electromagnetic environment specified below. The customer or user of
the REF 880 and 890 Series Video Colposcope should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
Electromagnetic Emissions
Electromagnetic Environment – Guidance
The REF 880 and 890 Series Video Colposcope uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The REF 880 and 890 Series Video Colposcope is suitable for use in all establishments,
including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic purposes.