Connections and Assembly

Connecting Video Colposcope with Monitor

WARNING If peripheral devices (example: monitor, video printer, VCR ,other) do
not comply with IEC 60601-1-1, they must be kept out of the patient area (6 feet
minimum from patient).
WARNING All signal input and output (I/O) connectors are intended for
connection to only peripheral devices (example: monitor, video printer, VCR)
that are in compliance with IEC 60601-1 (General Requirements for Safety,
Medical Electrical Equipment) or other IEC standards (for example IEC 60950,
Information Technology Equipment – Safety) as appropriate to the nature of the
peripheral device. Connecting additional peripherals to the Video Colposcope
may increase the risk associated with chassis or patient leakage currents.
To maintain operator and patient safety, the User should consider the system
leakage current requirements of IEC 60601-1-1 (Medical Electrical Equipment,
Safety Requirements for Medical Electrical Systems). The user should measure
leakage currents accordingly to confirm that no electric shock hazard exists.
An isolation transformer that is in compliance with IEC 60601-1 used to power
the additional peripherals may be used to control the system leakage current to
comply with the requirements of IEC 60601-1-1.
S-Video Cable to Monitor
Line A S-Video (Y/C) In
Back Panel of Monitor
LINE A
PARALLEL REMOTE RGB/COMPONENT
IN
OUT
IN
OUT
S-VIDEO
G/Y
LINE B
IN
IN
IN
OUT
OUT
OUT
VIDEO
VIDEO
B/Pb
~AC IN
IN
OUT
IN
IN
OUT
IN
OUT
OUT
AUDIO
AUDIO
R/Pr
AUDIO
17