6. Post-use Maintenance

After each patient maintain the product as follows.
WARNING
• Immediately after a treatment, the handpiece
should be cleaned, lubricated and sterilized.
Failure to properly maintain the handpiece may
lead to overheating, causing burn injuries or
product failure.
CAUTION
• Do not use the following fluids to wipe, immerse or clean
the product; strong/super acid water, strong acid/alkaline
chemicals, chlorine-containing solutions, solvents such
as benzine or thinner.
4
6-1 Preparation
1) Wear a eye protection, a mask and gloves to avoid
infection.
2) Remove the Rubber Cup / Brush.
3) Disconnect the handpiece from the motor.